Functional Features
NR20XE Large Aperture Precision Colorimeter
When testing products such as meat by ordinary colorimeter, the whole instrument body has to be placed in the object which not only stains the instrument, but also will let the liquid/water enter the instrument directly and bring cross contamination. However, NR20XE is a vertical structure. When measuring, it can make sure only the aperture touches the object which will not stain the instrument body.
NR20XE can meet quality control requirements in plastic (PET,PE,PP,PTA,ABS,EVA particles and powder), color steel plate, coating, textile, fluorescent material, denim, medicine and food industries, etc. It can help to control product color difference, improve product competitiveness, reduce enterprises’ loss and control production cost.
Food, medicine, textile, non-woven fabrics, denim, dyeing and printing, color steel plate, paper, plastic, chemical, coating, ink, building material industries, etc.
3nh
|
NR20XE Precision Colorimeter
|
Illuminating/Viewing Geometry
|
45°/0°
|
Measuring Aperture
|
Φ20mm
|
Special Function
|
Especially suitable for testing surface stripe and wet products
|
Locating
|
Illumination Locating/Cross Locating
|
Detector
|
Silicon photoelectric diode
|
Locating
|
Illuminating Locating/Cross Locating
|
Color Space
|
CIEL*a*b*C*h*; CIEL*a*b*; CIEXYZ
|
Color Difference Formula
|
△E*ab; △L*a*b*; △E*C*h*
|
Light Source
|
D65
|
Light Source Device
|
LED blue light excitation
|
Errors Between Each
Equipment |
≤0.50ΔE*ab
|
Storage
|
100pcs standards 20000pcs samples
|
Repeatability
|
Standard deviation within ΔE*ab 0.08 Average of 30 measurements of standard white plate
|
Weight
|
500g
|
Dimension
|
205×67×80 mm
|
Power source
|
Rechargeable lithium-ion battery 3.7V@3200mAh
|
Lamp Life
|
5 years, more than 1.6 million measurements
|
Charging Time
|
First charging time is 8 hours--100% electricity
|
PC Software
|
CQCS3 Software
|
Printer (optional)
|
Miniature thermal printer
|
ProductAdvantages
1. 20mm measuring aperture can test objects with surface stripe accurately.
2. Vertical structure can test wet surface object or can be soaked into the liquid for testing.
3. Double Locating: Illuminating locating and cross locating.
4. Adopting new optical path and dynamic integration time can test more stable and accurate.
5. 45°/0° optical geometry structure, considering the effects of surface features.
7. Equipped with rechargeable high-capacity Li-ion battery. No need to purchase battery repeatedly.
8. Configuring CQCS3 software -- connect PC computer to realize more functions.